Vacuum variable-angle far-infrared ellipsometer

Warning

This publication doesn't include Faculty of Education. It includes Faculty of Science. Official publication website can be found on muni.cz.
Authors

FRIŠ Pavel DUBROKA Adam

Year of publication 2017
Type Article in Periodical
Magazine / Source Applied Surface Science
MU Faculty or unit

Faculty of Science

Citation
Doi http://dx.doi.org/10.1016/j.apsusc.2016.10.125
Keywords Far infrared ellipsometry; Phonons; SrTiO3;
Description We present the design and performance of a vacuum far-infrared (similar to 50-680 cm(-1)) ellipsometer witha rotating analyser. The system is based on a Fourier transform spectrometer, an in-house built ellipsometer chamber and a closed-cycle bolometer. The ellipsometer chamber is equipped with a computer controlled theta-2 theta goniometer for automated measurements at various angles of incidence. We compare our measurements on SrTiO3 crystal with the results acquired above 300 cm(-1) with a commercially available ellipsometer system. After the calibration of the angle of incidence and after taking into account the finite reflectivity of mirrors in the detector part we obtain a very good agreement between the data from the two instruments. The system can be supplemented with a closed-cycle He cryostat for measurements between 5 and 400 K. (C) 2016 Published by Elsevier B.V.