Microfluidic device integrating single-cell extraction and electrical lysis for mass spectrometry detection of intracellular compounds

Investor logo

Warning

This publication doesn't include Faculty of Education. It includes Faculty of Science. Official publication website can be found on muni.cz.
Authors

VÁCLAVEK Tomáš FORET František

Year of publication 2023
Type Article in Periodical
Magazine / Source Electrophoresis
MU Faculty or unit

Faculty of Science

Citation
Web https://analyticalsciencejournals.onlinelibrary.wiley.com/doi/full/10.1002/elps.202100379
Doi http://dx.doi.org/10.1002/elps.202100379
Keywords bioanalysis; hemoglobin; mass spectrometry; microfluidics; single-cell analysis
Description Analysis of cellular composition and metabolism at a single-cell resolution allows gaining more information about complex relationships of cells within tissues or whole living organisms by resolving the variance stemming from the cellular heterogeneity. Mass spectrometry (MS) is a perfect analytical tool satisfying the demanding requirements of detecting and identifying compounds present in such ultralow-volume samples of high chemical complexity. However, the method of sampling and sample ionization is crucial in obtaining relevant information. In this work, we present a microfluidic sampling platform that integrates single-cell extraction from MS-incompatible media with electrical cell lysis and nanoESI-MS analysis of human erythrocytes. Hemoglobin alpha and beta chains (300 amol/cell) were successfully identified in mass spectra of single-erythrocyte lysates.