Comparative Study of RF Discharge in Neon by OES, Langmuir Probe, CR and PIC/MC Modelling

Investor logo

Warning

This publication doesn't include Faculty of Education. It includes Faculty of Science. Official publication website can be found on muni.cz.
Authors

NAVRÁTIL Zdeněk DVOŘÁK Pavel BRZOBOHATÝ Oto TRUNEC David

Year of publication 2010
Type Conference abstract
MU Faculty or unit

Faculty of Science

Citation
Description A novel method of determination of electron temperature and electron density in plasma based on optical emission spectroscopy and collisional-radiative modelling was compared with Langmuir probe measurement and PIC/MC simulation. An RF(13.56 MHz) capacitively-coupled discharge in neon at 10 Pa was investigated by intensity-calibrated optical emission spectroscopy. The intensities of neon transitions between 3s-3p states were fitted with CR model to determine the electron temperature and density. Metastable density measured in absorption was used in the calculations. A good agreement between electron temperature measured by OES+CR method and PIC/MC simulation was obtained close to driven electrode. The decrease of the electron temperature going into the bulk plasma, as expected from PIC/MC simulation, was not observed in experiment. Electron density determined from the fit of absolute intensities was found to be 5$\times$ larger than the result of Langmuir probe measurement. The axial dependencies of electron density determined by OES+CR and Langmuir probe were very similar.
Related projects:

You are running an old browser version. We recommend updating your browser to its latest version.